CS510 OXYGUARD DISSOLVED OXYGEN PROBE
INSTRUCTION MANUAL
1/00
COPYRIGHT (c) 2000 CAMPBELL SCIENTIFIC, INC.
Page view 0
1 2 3 4 5 6 ... 11 12

Summary of Contents

Page 1 - INSTRUCTION MANUAL

CS510 OXYGUARD DISSOLVED OXYGEN PROBEINSTRUCTION MANUAL1/00COPYRIGHT (c) 2000 CAMPBELL SCIENTIFIC, INC.

Page 2

CS510 OxyGuard Dissolved Oxygen Probe63) Determine the air temperature and barometric pressure.4) Using a calibration chart such as that provided in

Page 3

CS510 OxyGuard Dissolved Oxygen Probe7The following instructions in program table 2 would trigger the agitator asdiscussed in the agitator manual:*Tab

Page 4

CS510 OxyGuard Dissolved Oxygen Probe8This is a blank page.

Page 5 - Oxygen Probe

This is a blank page.

Page 6 - 2. Specifications

The CS510-L Oxyguard Dissolved Oxygen Probe is manufactured by Point Four SystemsIncorporated. Point Four’s conditions of sale and return policy appl

Page 7 - 5. Optional Agitator

This is a blank page.

Page 8

1CS510 OxyGuard DissolvedOxygen ProbeAn accurate and reliable sensor is a critical element in any measurement system. TheOxyGuard Stationary Probe mee

Page 9 - 9. Calibration

CS510 OxyGuard Dissolved Oxygen Probe22. SpecificationsPrinciple of Measurement: Membrane covered polarographic oxygen probe—self polarizing galvani

Page 10 - 11. Agitator Control

CS510 OxyGuard Dissolved Oxygen Probe33. Service KitSpare membrane cap with fitted membrane.50 ml electrolyte.Spare anode and O-ring.FIGURE 3-1. Exp

Page 11

CS510 OxyGuard Dissolved Oxygen Probe4FIGURE 5-1. Preventing Bio-fouling of the CS510-LThe PT4 Agitator overcomes these problems. The device is desig

Page 12 - This is a blank page

CS510 OxyGuard Dissolved Oxygen Probe5Optional Repeat Cycle Timer for Agitators: Reciprocating action may becontrolled by the optional solid state Rep